Photoelectric Reflectance Photometer ELREPHORemove the three screws (M 8><20) from the column and secure the latter to the
base plate using these screws. Tighten the screws. The column is correctly
positioned by the two pins (5 m 6x 25). Carefully place the measuring head on the
column and lock the appropriate socket head cap screw (M 16x70) in position
using the socket wrench supplied with the equipment. Do not yet tighten the socket
head cap screw. Securely screw the sample holder onto the base plate.
The ELREPHO operates on 220 V AC. 50 to 60 Hz. In case of other AC voltages.
a transformer is needed. which can be supplied by us. If no earthing contact is
available. connect the screw on the bottom of the power supply (adjacent to the
rubber foot) to a water pipe. radiator. or equivalent.
To correctly position the measuring aperture in relation to the sample holder.
connect the power supply to the mains and the ELREPHO. By switching on the
three toggle switches at the front of the power supply the lamps in the photometer
sphere light up.
The position of the measuring aperture in relation to the sample holder is seen
through the inspection window. Measuring aperture and sample holder must be
aligned to a common center.